CAMLOG ROOT-LINE 2 Surgical Procedure

15 CAMLOG ® ROOT-LINE 2 IMPLANT Planning template with CT-tubes for CT/DVT planning Template without upper tube section for use as drilling template X-ray template, outlined with tubes X-ray template with radio-opaque teeth, pre-inserted tubes and reference element for computer-based implant planning Ø 2.5 mm external diameter 10 mm 4 mm Example: CT-tubes for CT planning for pilot drill Ø 2.0 mm Drill for placement of CT-tubes Ø 2.1 mm internal diameter X-RAY/DRILLING TEMPLATE WITH CT-TUBES FOR CT PLANNING CT-tubes for the CT planning are integrated at the ideal implant positions in the planning templates created from the wax-up/set-up and are used as reference positions in the X-ray image. The CT-tubes have two parts, and the titanium material does not cause any scattering of rays in the CT/DVT. The lower section is polymerized into the template. The upper section is pluggable. The entire CT-tube is used for the radiological diagnostics; the upper section can be removed for surgery. Depending on the software used for the evaluation, titanium CT-tubes or other radio-opaque positioning elements (e.g. steel, barium sulfate) are integrated for the CT/DVT-sup- ported planning. Placing the CT-tubes directly on the mucosa makes it pos- sible to determine density in the CT/DVT. The documentation included with these systems contains more information on this topic.