CAMLOG ROOT-LINE 2 Surgical Procedure

22 CAMLOG ® ROOT-LINE 2 IMPLANT IMPLANT BED PREPARATION Overview: • Punch/mark the desired implant position with the Ø 2.3 mm round bur • Deep drill along the implant axial line with the Ø 2.0 mm pilot drill • Check with the Ø 1.7–2.8/2.0 mm paralleling pin with depth marks • Pre-drill with the Ø 1.7–2.8 mm pre-drill • Check with the Ø 1.7–2.8/2.0 mm paralleling pin with depth marks • Shape with the form drill • Probe the implant bed hole for its bony end • Tap ROOT-LINE 2 (for bone qualities 1 and 2*) We recommend using the tap for bone qualities 1* and 2*. 1.0 mm 1.0 mm Promote ® plus 0.4 mm Optional for all diameters Ø 3.3 mm Round bur Ø 2.3 mm rpm 800 800 16 mm 13 mm 11 mm Pilot drill Ø 2.0 mm Depth stop U/min 800 800 Ø 3.8 mm Ø 4.3 mm Ø 5.0 mm Ø 6.0 mm Round bur Ø 2.3 mm Round bur Ø 2.3 mm Round bur Ø 2.3 mm Round bur Ø 2.3 mm rp rpm 800 800 rpm rpm 800 800 800 800 16 mm 13 mm 11 mm 9 mm Pilot drill Ø 2.0 mm 16 mm 13 mm 11 mm 9 mm Pilot drill Ø 2.0 mm 16 mm 16 mm 13 mm 13 mm 11 mm 11 mm 9 mm 9 mm Pilot drill Ø 2.0 mm Pilot drill Ø 2.0 mm *Bone quality as documented in Lekholm & Zarb, 1985 Depth stop 9 mm Paralleling pin Depth measurement Paralleling pin Depth measurement Paralleling pin Depth measurement Paralleling pin Depth measurement Paralleling pin Depth measurement Depth stop Depth stop Depth stop IMPLANT POSITION PLANNING OVERVIEW OF DRILL SEQUENCES