CAMLOG ROOT-LINE 2 Surgical Procedure

25 CAMLOG ® ROOT-LINE 2 IMPLANT Shaping the implant bed includes cortical bone marking, pilot drilling, pre- drilling, and form drilling. The pilot- and pre-drill hole define the depth and axis of the implant bed and ensure conservative bone preparation by enlarging the diameter in small gradations. DRILL EXTENSION A drill extension is available to prevent resting of the angled handpiece on the remaining dentition during preparation of the implant bed adjacent to elon- gated teeth. DEPTH STOP The pilot drill SCREW-LINE*, reduced coil, and pre-drill SCREW-LINE* have a maximum working length of 16 mm. The drilling depths of 7, 9, 11, and 13 mm are laser-marked. Insertable depth stops limit the drilling depths to the selected depths of 9, 11, or 13 mm. CAUTION! The pilot drills and pre-drills SCREW-LINE* have a reduced diameter at the drill coil. The depth stops SCREW-LINE* are compatible only with these drills. Drill extension Depth stop SCREW-LINE* * also for ROOT-LINE 2 IMPLANT BED PREPARATION