CAMLOG ROOT-LINE 2 Surgical Procedure
34 CAMLOG ® ROOT-LINE 2 IMPLANT IMPLANT POSITIONING Protocol-compliant insertion depth of CAMLOG ® ROOT-LINE 2 implants with attached depth stop is achieved when the machined plane surface of the implant extends 0.4 mm above the crestal bone level and one of the three groove marks indicates the specified prosthetic position. If it was decided to set preparation depths for the implants individually by removing the depth stop during form drilling, this must be kept in mind when inserting the implant. It is possible to individually position implants vertically to match the drilling depth. GROOVE POSITIONING Corresponding to the three grooves of the CAMLOG implant-abutment connection, marks are inscribed on the drivers for the CAMLOG Implants. These permit a check of the groove positions during the insertion and their orientation as required for the prosthesis. If the dental technician has not indicated the groove position, a vestibular orientation is advantageous in most cases since the angle of angulated abutments originates at a groove. Note: Keep in mind during positioning of the grooves that turning to the next groove position (120°) will cause the screw implant to be inserted about 0.2 mm deeper. Driver CAMLOG ® insertion post CAMLOG ® ROOT-LINE 2 implant Groove marking Cam Groove 0.4 mm Vestibular orientation of the groove SURGICAL PROCEDURE