CAMLOG ROOT-LINE 2 Surgical Procedure
35 CAMLOG ® ROOT-LINE 2 IMPLANT Inserting the implant IMPLANT INSERTION The CAMLOG ® ROOT-LINE 2 implant is taken by the sterile handle and removed from the primary package (blister). Contamination from nonster- ile parts must be avoided. CAUTION Before inserting the implant, the silicone plug and the CAMLOG ® cover screw must be removed from the handle. Taking the implant by the handle, insert it into the implant bed. Make sure to observe the axial orientation of the implant bed. If a hole has been pre- tapped, the positions of the thread dog point in the cortical bone must match that on the implant. TIP: Carefully screw the implant to the left until you can feel the thread dog point. Then, using the handle, screw the implant to the right until it obtains sufficient grip to allow withdrawal of the handle. OPTIONAL ACCESSORIES: PICKUP INSTRUMENT If the intraoral space is insufficient to allow insertion of the implant with the handle, the implant can be removed from the handle using the PickUp instrument. Slide the PickUp instrument between the implant and handle onto the CAMLOG ® insertion post and lift off the handle. For the insertion procedure, place the selected driver on the CAMLOG ® insertion post. The implant is inserted into the bone, and the PickUp instrument is removed. IMPORTANT NOTE The instruments used in the CAMLOG ® Implant System are supplied non-sterile unless they are explicitly marked as sterile. They must be cleaned, disinfected and sterilized before first use and every use. IMPLANTATION