CAMLOG ROOT-LINE 2 Surgical Procedure

39 CAMLOG ® ROOT-LINE 2 IMPLANT SUBMERGED HEALING Using the screwdriver, hex, pick up the CAMLOG ® cover screw you removed from the handle and manually screw it into the CAMLOG ® ROOT-LINE 2 implant (Danger of aspiration!). The cover screw must only be tight- ened manually using the hex screwdriver. OPEN/TRANSGINGIVAL HEALING In this example, a cylindrical CAMLOG ® healing cap is picked up with the screwdriver, hex, and manually screwed into the CAMLOG ® ROOT-LINE 2 implant (Danger of aspiration!). The CAMLOG ® healing cap must be tightened only by hand with the screwdriver, hex. Inserting the CAMLOG ® healing cap CAMLOG ® ROOT-LINE 2 implant with healing cap Wound closure Inserting the CAMLOG ® cover screw CAMLOG ® ROOT-LINE 2 implant with cover screw Wound closure