CAMLOG ROOT-LINE 2 Surgical Procedure

40 CAMLOG ® ROOT-LINE 2 IMPLANT CAMLOG ® ADAPTER FOR SCREW IMPLANTS FOR NARROW GAPS AND HOLDING SLEEVE TO GUIDE THE ADAPTER INTO THE IMPLANT Diameter-reduced CAMLOG ® adapters, long, are available for the CAMLOG ® ROOT-LINE 2 implant diameters Ø 3.3, 3.8 mm and 4.3 mm for manual insertion into small gaps. Three lateral marks match the cams on the adapter and provide a buccal orientation for one of the CAMLOG ® ROOT-LINE 2 Implant grooves. After removal from the primary package (blister) the color-coded holding sleeve with correct diameter is inserted over the endosseous part of the implant to guide the adapter into the im- plant. The sleeve is squeezed together like a collet at the level of the implant shoulder and the handle is removed. NOTE In order to be able to hold the implant tightly and securely in the sleeve, we recommend clamping the sleeve using an hemostatic clip. Then, the handle can be removed, the fixing screw of the insertion post can be unscrewed using a screwdriver, hex, and the insertion post can be re- moved. IMPORTANT NOTE The hemostatic clip, the CAMLOG ® adapter and the sleeve must be sterilized prior to use. ACCESSORIES Setting up the sleeve Clamp the sleeve in an hemostatic clip and remove the handle Unscrewing the fixing screw Removing the CAMLOG ® insertion post CAMLOG ® adapter for screw implants, long, Ø 3.3 mm, Ø 3.8 mm and Ø 4.3 mm Holding sleeve for screw implants to guide the adapter into the screw implant, Ø 3.3 mm, Ø 3.8 mm and Ø 4.3 mm SURGICAL PROCEDURE