CAMLOG ROOT-LINE 2 Surgical Procedure
54 CAMLOG ® ROOT-LINE 2 IMPLANT NOTE: THE FIXING SCREW PROTRUDES ABOUT 2 MM FROM THE POST AFTER THE IMPRESSION POST ENGAGES AND BEFORE TIGHTENING. The fixing screw must be hand-tightened with the screwdriver, hex. For tight and thick gingiva in particular, we recommend a radiographic check of the correct seating of the impression post prior to taking the impression. IMPRESSION-TAKING The color-coded impression cap is now installed, using the guide grooves on the impression post, until a detectable pressure point is reached and the impression cap is definitely fastened. Three guide grooves on the impres- sion post (placed at 120° intervals) facilitate contact-free placement rela- tive to adjacent impression caps or teeth. The impression cap extensions must not be removed. Correct seating of the impression caps should be checked again before the impression is taken. The impression caps should remain in the impression after the impression tray is lifted. If this is not the case, repeat the impression-taking. Tip: To simplify the procedure, we recommend also sending the matching lab analog to the laboratory. To prevent loss of the fixing screw, the impression post must be shipped attached to the lab analog. Three guide grooves make three positioning options possible for the impression cap 1 3 2 IMPRESSION TAKING